| Model |
eQ mini-A |
eQ mini-B |
eQ mini-C |
| Sample volume |
16*0.2ml single tube or 8-strip transparent tube(except for white opaque tube) , you can mark the tube with markers on the cover of the tube. |
| Applicable reagents |
The system is an open platform, suitable for various real time PCR reagents, including fast extraction reagents, non-extract on reagents or direct amp l fat on reagents and other low-concentra to PCR reagents. |
| Reaction volume |
5-100ul |
| Dynamic range |
1-10'°Copies |
| Color wavelength |
400-700nm |
| Dyes and probes |
CH1: FAM、 SYBR Green l;
CH2:HEX、JOE、VIC、ROX. |
CH1: FAM、SYBR Green l;
CH2: HEX、JOE、VIC、TAMRA;
CH3: ROX、CALRED;
CH4: CY5、QUASAR. |
| Excitation light source |
High brightness and long life single color LED module. |
| Optical sensor |
PMT |
PD |
PD |
| Detection technology |
Scientific grade, suitable for
micro-reaction system. |
Double PD detection,practical grade,
suitable for conventional reaction
system. |
Four PD detection,independent
excitation fluorescence detection,
suitable for multiple PCR. |
| Channel number |
2-fluorescent channel |
4-fluorescent channel |
| ★Unique design of DFD photoelectric module |
(①DFD two-dimensional precision optical module, independent excitation light for each channel, independent fluorescence
detection, strong signal, low background noise. lt is especially suitable for fluorescent quantitative PCR experiments with
low-concentration DNA reagents such as direct amplification reagents, extraction-free reagents, and quick extraction reagents.
2All channels are detected in one scan, and each channel only needs 1 second. |
| Sensitivity |
1 copy |
| Rn |
≥0.998 |
| CV |
CV≤0.6% |
| Block setting range |
10-99.9℃ |
| Temperature control mode |
Tube mode.Real temperature of simulated reagent. |
| Refrigeration technology |
Peltier |
| ★ETSC block temperature
equalization technology |
The know-how of block edge temperature self-compensation design (ETSC) not only reduces the weight of the block, but also
achieves excellent temperature uniformity, which effectively shortens the time of PCR experiment. |
| Precision of temperature
control |
≤0.1℃ |
| Display resolution |
0.1°C |
| Temperature accuracy |
≤0.2℃ |
| Temperature uniformity |
<±0.2°C |
| Max.ramping rate |
≥4°C /s |
| Avg.ramping rate |
≥2.5°C /s (50~90°C) |
| Hot lid |
30-105°C,Default 105°C. |
| Operating mode |
Touchscreen control, with software. |
| Com.interface |
RS 232./USB |
| Software functions |
it has system parameter setting (with password authority), experimental parameter setting, sample information input.
operation management, data export EXCEL, PCR program overview, real-time amplification curve display, channel crosstalk
correction, criterion setting, automatic report, curve capture and other functions |
| Analytical functions
|
Standard Curve, Relative Quantitative Curve and Melting Curve (HRM). |
| Operating indicator
|
Power on, in operation, successful communication, ultraviolet degradation (UV), fault alarm, hot lid open alarm. |
| operating ambient |
Ambient temperature:10-30C,relative humidity:20-85%RH, elevation not higher than 2000 m, storage temperature:-20-60℃. |
| input power
|
100-240V~300VA |
| Dimensions (L*W*H) |
265x334x172mm |
| Net weight |
6.8kg |