EXPEC 4820 |
EXPEC 4850 |
| Optical System |
Paschen-Runge optical system using multiple high-performance CMOS detectors |
| |
Optical chamber with argon self-purification system |
Dual chamber argon purging optical system |
| |
Wavelength Range: 174~520 nm |
Wavelength Range:140~680nm |
| Source |
Programmable full digital pulse source |
| |
High energy pre-ignition technology, ignition pulse: 1~14Kv |
| |
Maximum discharge frequency: 1000Hz, Maximum discharge current: 400A |
| |
Spark excitation pulse: 20~230V, Arc excitation pulse: 20~60 V |
| Power supply and environmental requirements |
Power supply voltage: (220 ± 20) V AC, (50 ± 1) Hz, single power supply with protective grounding |
| |
Maximum excitation frequency: 400W, Average standby power: 50W |
| |
Working temperature: 10~30 ℃, Storage temperature: 0~45 ℃, Working humidity: 20~80% RH |
| Spark Stand |
Opened sample operation platform, which can be equipped with various sample adapters
for special applications to meet different small sample analysis needs |
| |
Optimized argon flushing design, better argon saving and spark stability |
| |
Easy to clean and maintain, intelligent reminder of maintenance cycle |
| Argon |
Aperture of spark stand: 13mm (optional: 6mm) |
| |
Purity: 99.999%, Pressure: 0.5MPa |
| |
Burning flow is about 3.5L/min, maintaining
and standby flow is about 0.1L/min |
Burning flow is about 3.5L/min, maintaining
and standby flow is about 0.2L/min |
| Dimensions and
weight |
Length: 818mm, width: 590, height: 396mm
weight: about 70Kg |
Length: 623mm, width: 735mm, height: 443mm
weight: about 80Kg |
| EMC |
IEC 610004-2, 61000-4-4, 61000-4-5 |
| Model |
F |
None |
Dual optical chamber configuration, best UV
analysis capability, capable of accurately
analyzing N elements, meeting high-end analysis
requirements for various matrixes such as Fe, Al,
Cu, Zn, Ni, Ti, Mg, Co, etc. |
|
N |
Suitable for all matrix analysis and can meet the matrix analysis requirements such as Fe,Al,Cu, Ni,and Zn |
Dual optical chamber configuration, strong UV analysis capability, meeting various matrix analysis requirements for Fe, Al, Cu, Zn, Ni, Ti, Mg,Co, etc. |
|
S |
Suitable for nonferrous metal analysis and meet the matrix analysis requirements such as Al,Cu, Zn, and Mg |
Single optical chamber configuration to meet matrix analysis requirements for Al, Cu, Zn, Mg,etc. |