Overview
Key Advantages
- High-efficiency EI ion source with modularized design for increased ionization yield
- Turbo molecular pump ensures stability and rapid vacuum recovery
- Electronic pressure control (EPC) improves retention-time reproducibility and injection consistency
- High-energy dynode electron multiplier provides enhanced sensitivity for trace-level detection
- Supports full-scan, SIM, TIC, and MC monitoring for flexible analytical workflows
- Integrated ChemAnalyst software enables data acquisition, method control, and automatic tuning
- Low operating cost with self-developed consumables and 24-hour technical support availability
Model Specifications
| Model |
Specifications |
| GC-MS 6800 |
450°C inlet, 0^100 psi EPC control, 120°C/min heating, EI 5–250 eV, 1.5–1000 amu range, 1000 amu/s scan rate, turbo pump 67 L/s |
Testing Applications
The GC-MS 6800 provides qualitative and quantitative analysis for industrial, pharmaceutical, environmental, and forensic applications.
FAQ
What inlet and oven temperatures does the GC-MS 6800 support?
The inlet temperature is up to 450 °C. The column oven works from room temperature +4 °C to 450 °C, with heating rates up to 120 °C/min and temperature programming of 7 stages / 8 platforms.
How is carrier gas pressure controlled?
The GC-MS 6800 uses electronic pressure control (EPC) with constant-velocity (CV) and constant-current (CC) support. EPC improves retention-time reproducibility and injection consistency.
What mass range and scan rate does it offer?
The mass range is 1.5-1000 amu with unit resolution (full width at half maximum) and a scan rate of up to 1000 amu/s.
What is the sensitivity of the GC-MS 6800?
In full scan mode the sensitivity is 1 pg OFN at m/z 272 with S/N of 30:1 or better (RMS). A high-energy dynode electron multiplier supports trace-level detection.
Which acquisition modes are supported?
The system supports full-scan, SIM, TIC, and MC monitoring.
What vacuum system does the GC-MS 6800 use?
A turbo molecular pump (67 L/s) provides stable vacuum and rapid vacuum recovery.
Which software controls the instrument?
The integrated ChemAnalyst software handles data acquisition, method control, and automatic tuning.
Can an autosampler be added?
Yes. An auto sampler is available as an option. The GC also supports split/splitless injection with a maximum split ratio of 1000:1.