EDS9800T Precious Metal Analyzer for XRF Elemental Quantification

EDS9800T Precious Metal Analyzer for XRF Elemental Quantification

Model :EDS9800T

The AELAB EDS9800T Precious Metal Analyzer is a benchtop X-ray fluorescence spectrum analyzer engineered for fast, accurate elemental quantification in precious metal testing. Designed for solid, powder, and liquid samples, the EDS9800T analyzes composition from ppm to 99.99% with 0.01% accuracy (Au≥96%). Featuring an SDD detector, digital multichannel analyzer, and qualitative and quantitative analysis software, it supports rapid multi-element evaluation for jewelry and alloy quality control.

Need Help with This Product?

Need Help with This Product?

+8615800016110

Overview

The AELAB EDS9800T Precious Metal Analyzer is a benchtop X – Fluorescence Spectrum Analyzer designed to analyzes precious metal composition and measures elemental concentration for quality control in jewelry, bullion, and alloy verification. With an analysis range from ppm to 99.99% and 0.01% accuracy (Au≥96%), the EDS9800T quantifies trace and major elements with high repeatability.

Built around an SDD detector and digital multichannel acquisition, the AELAB EDS9800T Precious Metal Analyzer stabilizes measurement performance through selectable customized filter switches and a 5 kV to 50 kV tube voltage configuration. Fast test cycles of 5 to 60 seconds support efficient precious metal screening without compromising analytical resolution.

 

⚙️ Design & Construction

The AELAB EDS9800T Precious Metal Analyzer integrates a compact benchtop enclosure with a controlled sample chamber sized 310 × 270 × 90 mm, supporting stable positioning of solid, powder, and liquid specimens. The system includes an HD camera for visual sample alignment and inspection, enabling precise placement during XRF testing workflows. The instrument dimensions of 480 × 440 × 368 mm and weight of 36 kg support laboratory and production-floor installation.

 

🔬 Working Principle / Testing Method

The AELAB EDS9800T Precious Metal Analyzer performs X-ray fluorescence spectrum analysis by irradiating a sample with an X-ray tube operating at 5 kV to 50 kV. Emitted characteristic fluorescence signals are detected by the SDD detector and processed by the digital multichannel analyzer to quantify elemental composition. The system analyzes 82 measurable elements from sodium (Na) to uranium (U), and supports up to 26 elements simultaneously per measurement cycle (for use in the precious metals industry only).

 

💡 Key Advantages

  • The AELAB EDS9800T Precious Metal Analyzer quantifies composition across ppm to 99.99% concentration range
  • Provides 0.01% accuracy (Au≥96%) for high-confidence precious metal verification
  • Uses an SDD detector with 135 ± 5 eV resolution to improves spectral discrimination
  • Delivers rapid measurement cycles with 5 to 60 seconds test time for high-throughput workflows
  • Supports 82 elements from sodium (Na) to uranium (U), enabling broad alloy characterization
  • Measures up to 26 elements simultaneously per test cycle for multi-element quantification
  • Includes selectable customized filter switches for optimized excitation and matrix compensation

 

🔗 Connectivity & Integration

The AELAB EDS9800T Precious Metal Analyzer includes qualitative and quantitative analysis software for elemental identification and concentration quantification. The software supports standardized analytical workflows suitable for laboratory documentation, quality control reporting, and traceability-driven operations. Integration options beyond the included software are not specified in the provided catalog data.

 

🧪 Testing Applications

The AELAB EDS9800T Precious Metal Analyzer is optimized for XRF elemental quantification in precious metal and alloy testing environments. Common applications include gold purity verification, jewelry alloy composition control, bullion material authentication, and multi-element screening where rapid decision-making is required. The EDS9800T supports analysis of solid, powder, and liquid samples, enabling flexible workflows across jewelry manufacturing, recycling, assay laboratories, and material trading operations.

 

Compliance & Quality Assurance

The AELAB EDS9800T Precious Metal Analyzer is designed for controlled laboratory and industrial testing environments, with recommended operating conditions of 5 °C to 38 °C and 15% to 90% relative humidity. The power requirement supports AC220 V ± 10 V or AC110 V ± 10 V at 50 / 60 Hz. To maintain measurement stability, installation should avoid high-power electromagnetic sources and vibration interference. Certification standards (ISO / CE / GLP) and calibration traceability methods are not specified in the provided catalog data and should be verified for formal compliance documentation.

For laboratories and production teams requiring fast, accurate XRF composition verification, the AELAB EDS9800T Precious Metal Analyzer delivers stable multi-element quantification from ppm to 99.99% using SDD detection and software-based analytical processing—supporting confident precious metal testing under controlled quality assurance conditions.

AELAB EDS9800T Precious Metal Analyzer catalog showing device, key features, applications, and technical specifications.

Specifications

Technical Specifications – AELAB EDS9800T Precious Metal Analyzer
Parameter EDS9800T
Instrument Type X - Fluorescence Spectrum Analyzer
Analysis Range ppm to 99.99%
Accuracy 0.01% (Au≥96%)
Sample Form Solid, powder, liquid
Tube Voltage 5kV ~ 50kV
High Voltage Power Supply 0 ~ 50 kV
Camera HD camera
Filters Selectable customized switches
Detector Type SDD
Resolution 135 ± 5 eV
Multichannel Analyzer Digital multichannel
Sample Chamber Size 310 × 270 × 90 (mm)
Test Time 5 ~ 60 sec
Measuring Element 82 elements from sodium (Na) to uranium (U) between all elements, Up to 26 elements simultaneously each time (For use in the precious metals industry only)
Analysis Software qualitative and quantitative analysis software
External Dimensions 480 × 440 × 368 (mm)
Weight 36 kg
Ambient Temperature 5 °C ~ 38 °C (Suggested that in air conditioning room)
Relative Humidity 15% ~ 90%
Power Requirements AC220V ± 10V, or AC110V ± 10V, 50 / 60 Hz
Installation Requirement No high power electromagnetic and vibration interference sources nearby

Operation & Workflow

Follow these steps to safely power on, initialize, prepare, and test samples for accurate and reliable XRF analysis using the EDS9800T Precious Metal Analyzer.

Step 1: Power On & Preheating

  • Connect the instrument to a stable power supply.
  • Turn on the power switch (PWR).
  • Allow the machine to preheat for 5 minutes before launching the software.
  • Ensure the spectrometer software is not running during preheating.

Step 2: Initialize the Instrument

  • Launch the XRF software from the desktop shortcut.
  • Place the calibration sample and click the Initialize button.
  • Confirm by clicking OK. Initialization completes in approximately 10–20 seconds.

Step 3: Sample Preparation & Placement

  • Place your sample (solid, powder, or liquid) on the sample holder, ensuring contact with the test film.
  • Select the appropriate template for your sample type.
  • Coated samples should have the coating removed for accurate results.

Step 4: Running a Test

  • Click the Start button in the software interface.
  • Optionally, enter a sample name or leave blank.
  • The instrument automatically measures and analyzes the sample.
  • Repeat the test 3–5 times for averaging to ensure accuracy.

Step 5: Saving & Exporting Results

  • After the test completes, click Save.
  • You will be prompted to open the report in Excel; click OK to view or Cancel to save only.
  • All test results are preserved in the designated save path for traceability.

Key Operational Tips

  • Maintain a clean, dry workspace and avoid dusty or excessively hot/cold environments.
  • Use a UPS or power stabilizer for consistent performance.
  • Do not move or service the instrument without authorization.
  • Warm up the instrument for 10–30 minutes if performing consecutive tests.

Maintenance & Care

General Maintenance Guidelines

  • Assign a trained operator; do not operate or move without permission.
  • Keep the workspace clean and dry; avoid dusty or polluted environments.
  • Maintain ambient temperature between 5°C–38°C and relative humidity 15%–90%.
  • Use a UPS or power stabilizer to ensure stable operation.

Cleaning & Sample Handling

  • Regularly clean the instrument exterior and sample chamber.
  • Always close the cover after testing to prevent dust accumulation.
  • Ensure samples are properly positioned on the sample holder and contact the test film correctly.

Safety & Operational Warnings

  • Do not open or repair the device without authorized support.
  • Avoid placing near high-power electromagnetic sources, motors, or welding equipment.
  • For consecutive tests, warm up the instrument for 10–30 minutes.
  • Cut off coatings on samples before testing; improper templates or coated samples can lead to inaccurate results.
  • Do not judge sample results from a single test; always perform 3–5 measurements and calculate the average.

Key Tips for Longevity

  • Avoid extremely hot, cold, or humid rooms.
  • Handle carefully when moving to prevent mechanical or electronic damage.
  • Follow proper shutdown procedures after testing to preserve device performance.

Request Quote: EDS9800T Precious Metal Analyzer for XRF Elemental Quantification

Frequently Asked Questions

1What applications is the AELAB EDS9800T Precious Metal Analyzer suitable for?
The AELAB EDS9800T Precious Metal Analyzer is used for XRF elemental composition testing in precious metals, including gold purity verification, alloy identification, and multi-element quantification.
2What sample types can the EDS9800T test?
The AELAB EDS9800T Precious Metal Analyzer supports solid, powder, and liquid sample forms.
3What is the accuracy of the AELAB EDS9800T for gold analysis?
The AELAB EDS9800T Precious Metal Analyzer provides 0.01% accuracy when Au is ≥96%.
4How fast is a typical test on the AELAB EDS9800T Precious Metal Analyzer?
The AELAB EDS9800T Precious Metal Analyzer performs testing within 5 to 60 seconds depending on the method and sample composition.
5What environmental conditions are recommended for operation?
The AELAB EDS9800T Precious Metal Analyzer operates reliably at 5 °C to 38 °C and 15% to 90% relative humidity, with installation away from strong electromagnetic and vibration interference sources.

Request Quote: EDS9800T Precious Metal Analyzer for XRF Elemental Quantification

(0/10)