info@aelabgroup.com
QR Code 1
QR Code 2
Automatic PMCC Flash Point Tester SYD-261D | AELAB

Automatic PMCC Flash Point Tester SYD-261D | AELAB

Model :SYD-261D

  • Test Standards: GB/T 261-2008 and ASTM D93 Pensky-Martens closed cup methods
  • Flash Point Range: Ambient to 230℃ with 0.1℃ accuracy
  • Heating Rates: Procedure A (5~6)℃/min; Procedure B (1~1.5)℃/min; Procedure C (3.0±0.5)℃/min
  • Stirring Rates: Procedure A/C (90~120)RPM; Procedure B (250±10)RPM
  • Control & Data: 8-inch IPS capacitive screen, Bluetooth, real-time curves and automatic atmospheric-pressure correction

Overview

Product Overview

The AELAB SYD-261D is an Automatic PMCC Flash Point Tester designed to determine the closed cup flash point of petroleum products using the Pensky-Martens method.

The instrument is designed and manufactured according to GB/T 261-2008 Determination of Flash Point – Pensky-Martens Closed Cup Method and ASTM D93 Standard Test Method for Flash Point by Pensky-Martens Closed Cup Tester.

The SYD-261D provides a flash point determination range from ambient temperature to 230℃ with 0.1℃ accuracy. It uses a 32-bit ARM processor and high-precision AD chip to record test data and parameters in real time and display them as curves on an 8-inch IPS high-definition capacitive screen.

Temperature rise, cover opening, ignition, flash detection and data printing are completed automatically, while the test arm rises and falls automatically. The instrument also measures atmospheric pressure automatically and corrects the flash point value accordingly.

Key Features

  • 8-Inch IPS Display: Uses an 8-inch IPS high-definition capacitive screen for instrument operation and data display.
  • 32-Bit ARM Processing: A 32-bit ARM processor and high-precision AD chip record test data and parameters in real time and display them as curves.
  • Bluetooth Interface: Bluetooth data transmission allows stored test data to be queried.
  • Remote Instrument Functions: The catalog specifies remote automatic upgrading, remote prediction, abnormal early warning, and evaluation and maintenance of instrument operating status.
  • Atmospheric Pressure Correction: Automatically measures atmospheric pressure and corrects the flash point value.
  • Automatic Test Operation: Temperature rise, cover opening, ignition, detection and data printing are completed automatically, and the test arm rises and falls automatically.
  • Automatic Ignition and Cooling: Uses electronic ignition with gas flame and automatic air cooling at the end of the test.

Applications

  • Petroleum Products: Determination of closed cup flash point using the Pensky-Martens method.
  • GB/T 261-2008 Testing: Closed cup flash point determination according to the referenced Chinese standard.
  • ASTM D93 Testing: Pensky-Martens closed cup flash point determination according to the referenced ASTM method.

Test Methods and Standards

  • GB/T 261-2008: Determination of Flash Point – Pensky-Martens Closed Cup Method
  • ASTM D93: Standard Test Method for Flash Point by Pensky-Martens Closed Cup Tester

Product Information

  • Brand: AELAB
  • Manufacturer: AELAB GUANGZHOU CO., LTD.
  • Product Type: Automatic PMCC Flash Point Tester
  • Model: SYD-261D
  • Test Method: Pensky-Martens closed cup method
  • Applicable Samples: Petroleum products
  • Flash Point Range: Ambient to 230℃
  • Flash Point Accuracy: 0.1℃
  • Data Transmission: Bluetooth

FAQ

What is the SYD-261D Automatic PMCC Flash Point Tester used for?

It is used to determine the closed cup flash point of petroleum products using the Pensky-Martens method.

Which standards does the SYD-261D follow?

The instrument is designed and manufactured according to GB/T 261-2008 and ASTM D93.

Is 261-2008 part of the SYD-261D model number?

No. The product model is SYD-261D. GB/T 261-2008 is one of the referenced test standards.

What is the flash point determination range and accuracy?

The specified range is from ambient temperature to 230℃ with an accuracy of 0.1℃.

What heating rates are available?

Procedure A uses (5~6)℃/min, Procedure B uses (1~1.5)℃/min, and Procedure C uses (3.0±0.5)℃/min.

What stirring rates are available?

Procedures A and C use (90~120)RPM, while Procedure B uses (250±10)RPM. Stirring is automatically controlled and can also be adjusted manually.

Does the SYD-261D correct for atmospheric pressure?

Yes. The instrument automatically measures atmospheric pressure and automatically corrects the flash point value.

Does the SYD-261D support Bluetooth?

Yes. The specified data transmission mode is Bluetooth.

Which parts of the test process are automatic?

Temperature rise, cover opening, ignition, detection and data printing are completed automatically. The test arm also rises and falls automatically, and automatic air cooling operates at the end of the test.

Who manufactures the SYD-261D?

The SYD-261D is supplied under the AELAB brand by AELAB GUANGZHOU CO., LTD.

Specifications

Parameter Specification
Power supply AC(220±5%)V,50Hz
Total power consumption ≤600W
Flash point determination Range: Ambient to 230℃ Accuracy: 0.1℃
Heating rate Procedure A: (5~6)℃/min Procedure B: (1~1.5)℃/min Procedure C: (3.0±0.5)℃/min
Stirring rate Procedure A and C: (90~120)RPM Procedure B: (250±10)RPM Automatic control and manually adjustable.
Igniting mode (1)Electric ignition. (2)Diameter of igniter:0.7mm~0.8mm
Data transmission mode Bluetooth
Suitable temperature Ambient temperature:(15~35)℃ Relative humidity:≤85%
Dimension 510mm×320mm×330mm
Net weight 20kg.

Request Quote: Automatic PMCC Flash Point Tester SYD-261D | AELAB

Request Quote: Automatic PMCC Flash Point Tester SYD-261D | AELAB